VPFlowScope M shows no Flow.
There can be two reasons for this:
1. If you have recently updated or going to update your VPFlowScope M to the latest firmware (2.2.0), please be careful about the diameter.
The diameter of the transmitter will be set to 0
after a firmware update. Even when the cartridge is reconnected, it remains at 0, and 0 is also put in the cartridge.
Consequently, The flow value is 0 until you set the correct diameter. You can solve this by entering the right diameter.
2. When powered from USB only, the flow sensor is not turned on. You need to power the instrument via 16...24 VDC (5 pin M12) as well.

VPFlowScope M measures pressure and temperature (no flow) while connecting only via USB. In order to read out the flow, you need to power up the instrument via the 5-pin port.
Related Articles
Standard checklist - flow meter support
This list of questions is essential to provide the right support. It helps us to be of quicker assistance, as we have a complete picture of the situation. Process: Inner Diameter of the pipe Upstream length Downstream length Average pressure in the ...
USB Isolator for connection to VPStudio
The VPFlowScope M can be connected to a computer and the VPStudio 4 software for adjusting settings or viewing real-time data. This connection can be made using either a JB5 kit or a USB Mini-B cable. However, when using a USB Mini-B cable, the power ...
RS485 Communication for VPFlowScope M
RS485 Communication Real time value of Temperature in Celsius (204-205) [RTU] Tx: 09 03 00 CC 00 02 05 7C Slave Address: 09 Function Code: 03 Starting Address: 00 CC Quantity of Registers: 00 02 CRC: 05 7C [RTU] Rx: 09 03 04 1C 20 41 CD 84 6C Slave ...
Compressed air flow meters - how to minimize installation errors
So you are measuring compressed air flow? Great! Now you can generate some serious savings. But wait! You have installed three meters downstream of a main flow meter, and you add them all up? and you see that the sum of these is not equal to the main ...
How to test communication through Modbus?
Using Modpoll freeware ( VP recommendation) Download it from the links below https://www.vpinstruments.com/service-support/software-firmware or https://www.modbusdriver.com/modpoll.html Unzip the file, open command line (cmd), and change the ...